Water And Soil Analysis Kit
Our water and soil analysis Kits are available as per clients’ specifications. Superior quality raw material and latest technology is used to fabricate these instruments. Our engineers and designers fabricate products which are top notch on account of long life, durability and available designs.They are Used For:
Measuring pH, ORP, Turbidity, conductivity, TDS, DO, Salinity & Temperature.
They are Further Classified Based upon your required Parameters :
Visen 71 K: MicroProcessor Analysis Kit. It is Used For measuring the pH , Temp. , D.O , TDS , Salinity , Conductivity , ORP .
pH | |
Range | 0 to 14pH |
Resolution | 0.01pH |
Accuracy | ±0.01pH (Relative) |
Temp. Comp | Automatic With Temp. Probe |
Buffer | Manual By Keys |
Recognition | 4.00 7.00, 9.20 pH (Automatic) |
Slope Adjustment | 80% to 120%, By Default 100% |
Calibration | Auto/Manual 2-Point, 3-Point |
mV (ORP) | |
Range | 0 to ±1999mV |
Resolution | 1mV |
Accuracy | ±1 |
mV Calibration | Manual |
Temperature | |
Range | -20.0 to +200°C |
Resolution | 0.1°C |
Accuracy | ±0.2% |
Input Type | Pt 100 |
Calibration | Manual |
Dissolved Oxygen | |
Range | 0 t 20pap |
Resolution | 0.1ppm |
Temperature Comp. | Automatic with pt 100probe & manual with keys |
Pressure Adjustment | 0 to 999mm of Hg Manual |
Salinity Correction | 0 to 20gm/1 (Manual) |
Calibration | Automatic |
Accuracy | ±2%±1digit |
Conductivity | |
Range | 0 to 1000m mho in 6-ranges |
Range Selection | Automatic 6 Ranges 0µ mho to 1000m mho |
Resolution | 0.1µ mho |
Cell Constant | Adjustable |
Temperature co.-eff | Adjustable between 0 to 4% |
Temperature Comp. | Automatic for entire range |
Accuracy | ±0.5% of range + 1digit |
TDS | |
Range | 0 to 200 ppt in 5 Ranges |
Range Selection | Automatic |
Resolution | 0.1 ppm |
TDS Factor | 0.5 and adjustable manuall |
SALINITY | |
Range | 0 to 50 ppt |
Resolution | 0.1 ppt |
Accuracy | ±0.5% of range + 1digit |
Temperature Comp. | Automatic |
Salinity Factor | 0.75 and Adjustable Manually |
FEATURES | |
Display | 16 Character * 2 lines Alphanumeric lcd. |
Display Format | Temp & pH or temp & ORP Simultaneously auto Display of Selected Buffer |
Setting | By soft Touch Keys |
Storage | 1000 samples With Date and Time |
Interface | RS-232 Centronics & Computer |
Real Time Display | 24 hour Mode With Date |
Error Indication | pH error , Buffer Error , ORP Error |
Digital Output |
|
Print Format |
|
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temeprature Probe 5) Conductivity/TDS/Salinity Cell 6) DO Probe 7) Instruction Manual |
Visen 91 : MicroProcessor Eight Parameters Analysis Kit. It is Used For measuring the pH , Temp. , D.O , TDS , Salinity , Turbidity , Conductivity , ORP .Water And Soil Analysis kit
pH | |
Range | 0 to 14pH |
Resolution | 0.01pH |
Accuracy | ±0.01pH (Relative) |
Temp. Comp | Automatic With Temp. Probe |
Buffer | Manual By Keys |
Recognition | 4.00 7.00, 9.20 pH (Automatic) |
Slope Adjustment | 80% to 120%, By Default 100% |
Calibration | Auto/Manual 2-Point, 3-Point |
mV (ORP) | |
Range | 0 to ±1999mV |
Resolution | 1mV |
Accuracy | ±1 |
mV Calibration | Manual |
Temperature | |
Range | -20.0 to +200°C |
Resolution | 0.1°C |
Accuracy | ±0.2% |
Input Type | Pt 100 |
Calibration | Manual |
Dissolved Oxygen | |
Range | 0 t 20pap |
Resolution | 0.1ppm |
Temperature Comp. | Automatic with pt 100probe & manual with keys |
Pressure Adjustment | 0 to 999mm of Hg Manual |
Salinity Correction | 0 to 20gm/1 (Manual) |
Calibration | Automatic |
Accuracy | ±2%±1digit |
Conductivity | |
Range | 0 to 1000m mho in 6-ranges |
Range Selection | Automatic 6 Ranges 0µ mho to 1000m mho |
Resolution | 0.1µ mho |
Cell Constant | Adjustable |
Temperature co.-eff | Adjustable between 0 to 4% |
Temperature Comp. | Automatic for entire range |
Accuracy | ±0.5% of range + 1digit |
Turbidity | |
Range | 0-100JTU |
Resolution | 1JTU |
Accuracy | ±3% FS ±1digit |
Sample Size | 30mm glass tube |
Light | 6.8V Tungsten Lamp |
TDS | |
Range | 0 to 200 ppt in 5 Ranges |
Range Selection | Automatic |
Resolution | 0.1 ppm |
TDS Factor | 0.5 and adjustable manuall |
SALINITY | |
Range | 0 to 50 ppt |
Resolution | 0.1 ppt |
Accuracy | ±0.5% of range + 1digit |
Temperature Comp. | Automatic |
Salinity Factor | 0.75 and Adjustable Manually |
FEATURES | |
Display | 16 Character * 2 lines Alphanumeric lcd. |
Display Format | Temp & pH or temp & ORP Simultaneously auto Display of Selected Buffer |
Setting | By soft Touch Keys |
Storage | 1000 samples With Date and Time |
Interface | RS-232 Centronics & Computer |
Real Time Display | 24 hour Mode With Date |
Error Indication | pH error , Buffer Error , ORP Error |
Digital Output |
|
Print Format |
|
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temeprature Probe 5) Conductivity/TDS/Salinity Cell 6) DO Probe 7) Instruction Manual |
Visen 991 K : 6T Parameter Kit. It is Used For measuring the pH , Temp. , D.O , TDS , Turbidity , Conductivity .
pH | |
Range | 0 to 14 pH |
Temp. Comp | 0-100°C |
Resolution | 0.01pH |
Accuracy | ±0.01pH ± 1 digit |
Conductivity | |
Range | 0-20ms, 0-200 |
Cell Constant | Pre-Adjustable |
Resolution | 0.1 µmhos |
Accuracy | ±05% FS ± 1 digit |
Temperature | |
Range | 0-100°C |
Resolution | 0.1 |
Accuracy | ±0.1°C ± 1 digit |
TDS | |
Range | 0-20ppt, 0-200ppt |
Cell Constant | Pre-Adjustable |
Resolution | 0.1ppm |
Accuracy | ±5% FDS ±1 digit |
D.O. | |
Range | 0-20ppm |
Resolution | 0.1ppm |
Accuracy | ±0.01C ±1digit |
Temp. Comp | 05°-50°C (Manual) |
Turbidity | |
Range | 0-100JTU |
Resolution | 1JTU |
Accuracy | ±3% FS ±1digit |
Sample Size | 30mm glass tube |
Light | 6.8V Tungsten Lamp |
Detector | Photodiode |
Display | 3 ½ digit LCD Display |
Power Supply | 6 dry cells of 1.5V OR 230±10PHz |
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) 3 Tube Set 4) Temperature Probe 5) Conductivity/TDS/Salinity Cell 6) DO Probe 7) Instruction Manual |
Visen 971 K : 7 Parameter Kit. It is Used For measuring the pH, Temp., D.O , TDS , ORP, Salinity, Conductivity .
pH | |
Range | 0 to 14 pH |
Temp. Comp | 0-100°C |
Resolution | 0.01pH |
Accuracy | ±0.01pH ± 1 digit |
Conductivity | |
Range | 0-20ms, 0-200 |
Cell Constant | Pre-Adjustable |
Resolution | 0.1 µmhos |
Accuracy | ±05% FS ± 1 digit |
Temperature | |
Range | 0-100°C |
Resolution | 0.1 |
Accuracy | ±0.1°C ± 1 digit |
TDS | |
Range | 0-20ppt, 0-200ppt |
Cell Constant | Pre-Adjustable |
Resolution | 0.1ppm |
Accuracy | ±5% FDS ±1 digit |
D.O. | |
Range | 0-20ppm |
Resolution | 0.1ppm |
Accuracy | ±0.01C ±1digit |
Temp. Comp | 05°-50°C (Manual) |
mV(ORP) | |
Range | 0 to ± 1999 mV |
Resolution | 1 mV |
Accuracy | ±1 |
mVCalibration | MANUAL |
Salinity | |
Range | 0-50 ppt |
Resolution | 0.1 ppt |
Temp. comp. | Automatic |
Accuracy | + 0.5% |
Salinity factor | 0.75 by default( adjustable) |
Display | 3 ½ digit LCD Display |
Power Supply | 6 dry cells of 1.5V OR 230±10PHz |
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temperature Probe 5) Conductivity/TDS/Salinity Cell 6) DO Probe 7) Instruction Manual |
Visen 972 K : 6 Parameter Kit. It is Used For measuring the pH, Temp., D.O , TDS , ORP, Conductivity .
pH | |
Range | 0 to 14 pH |
Temp. Comp | 0-100°C |
Resolution | 0.01pH |
Accuracy | ±0.01pH ± 1 digit |
Conductivity | |
Range | 0-20ms, 0-200 |
Cell Constant | Pre-Adjustable |
Resolution | 0.1 µmhos |
Accuracy | ±05% FS ± 1 digit |
Temperature | |
Range | 0-100°C |
Resolution | 0.1 |
Accuracy | ±0.1°C ± 1 digit |
TDS | |
Range | 0-20ppt, 0-200ppt |
Cell Constant | Pre-Adjustable |
Resolution | 0.1ppm |
Accuracy | ±5% FDS ±1 digit |
D.O. | |
Range | 0-20ppm |
Resolution | 0.1ppm |
Accuracy | ±0.01C ±1digit |
Temp. Comp | 05°-50°C (Manual) |
mV(ORP) | |
Range | 0 to ± 1999 mV |
Resolution | 1 mV |
Accuracy | ±1 |
mVCalibration | MANUAL |
Display | 3 ½ digit LCD Display |
Power Supply | 6 dry cells of 1.5V OR 230±10PHz |
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temperature Probe 5) Conductivity/TDS/Salinity Cell 6) DO Probe 7) Instruction Manual |
Visen 973 K : 5 Parameter Kit. It is Used For measuring the pH , Temp. , TDS , mv(ORP) , Conductivity .
pH | |
Range | 0 to 14 pH |
Temp. Comp | 0-100°C |
Resolution | 0.01pH |
Accuracy | ±0.01pH ± 1 digit |
Conductivity | |
Range | 0-20ms, 0-200 |
Cell Constant | Pre-Adjustable |
Resolution | 0.1 µmhos |
Accuracy | ±05% FS ± 1 digit |
Temperature | |
Range | 0-100°C |
Resolution | 0.1 |
Accuracy | ±0.1°C ± 1 digit |
TDS | |
Range | 0-20ppt, 0-200ppt |
Cell Constant | Pre-Adjustable |
Resolution | 0.1ppm |
Accuracy | ±5% FDS ±1 digit |
mV(ORP) | |
Range | 0 to ± 1999 mV |
Resolution | 1 mV |
Accuracy | ±1 |
mVCalibration | MANUAL |
Display | 3 ½ digit LCD Display |
Power Supply | 6 dry cells of 1.5V OR 230±10PHz |
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temperature Probe 5) Conductivity/TDS/Salinity Cell 6) Instruction Manual |
Visen 974 K : 4 Parameter Kit. It is Used For measuring the pH , Temp. , mv(ORP) , Conductivity .
pH | |
Range | 0 to 14 pH |
Temp. Comp | 0-100°C |
Resolution | 0.01pH |
Accuracy | ±0.01pH ± 1 digit |
Conductivity | |
Range | 0-20ms, 0-200 |
Cell Constant | Pre-Adjustable |
Resolution | 0.1 µmhos |
Accuracy | ±05% FS ± 1 digit |
Temperature | |
Range | 0-100°C |
Resolution | 0.1 |
Accuracy | ±0.1°C ± 1 digit |
mV(ORP) | |
Range | 0 to ± 1999 mV |
Resolution | 1 mV |
Accuracy | ±1 |
mVCalibration | MANUAL |
Display | 3 ½ digit LCD Display |
Power Supply | 6 dry cells of 1.5V OR 230±10PHz |
Accessories | 1) pH Electrode 2) Standard Solution 7.0 pH & 4.0 pH 3) ORP Electrode (On Demand) 4) Temperature Probe 5) Conductivity/TDS/Salinity Cell 6) Instruction Manual |